3-D position sensitive CdZnTe gamma-ray spectrometers
Two 3-D position-sensitive room temperature semiconductor γ-ray spectrometers have been built using 1 cm 3 cubic CdZnTe crystals. The lateral coordinates of γ-ray interaction are obtained from the location of the 11( x)×11( y) pixellated anodes and the depth ( z) is obtained from the ratio of the si...
Saved in:
Published in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Vol. 422; no. 1; pp. 173 - 178 |
---|---|
Main Authors: | , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
Elsevier B.V
1999
|
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Two 3-D position-sensitive room temperature semiconductor γ-ray spectrometers have been built using 1
cm
3 cubic CdZnTe crystals. The lateral coordinates of γ-ray interaction are obtained from the location of the 11(
x)×11(
y) pixellated anodes and the depth (
z) is obtained from the ratio of the signals coming from the cathode and the anode. Energy spectra from 662
keV incident γ-rays have been collected from each of the
11
(x)×11
(y)×20
(z)
voxels in both of the CdZnTe devices. After corrections for electron trapping, the difference of weighting potentials in 3-D, and for the gain variation of the readout circuitry, energy resolutions of 1.70% (11.3
keV) FWHM and 1.84% (12.2
keV) FWHM were obtained at 662
keV γ-ray energy on the first and second detectors, respectively, from the whole bulk for single-pixel events. Possible improvements in the detector performance are discussed. |
---|---|