A new heart-sounds gating device for medical imaging
A heart-sounds gating device has been designed and tested which identifies, individually, both the first (S1) and second (S2) heart sound from their timing relationship, providing two trigger points through the cardiac cycle for synchronizing medical images. The gate utilizes dynamically varying tim...
Saved in:
Published in: | IEEE transactions on biomedical engineering Vol. 39; no. 3; pp. 307 - 310 |
---|---|
Main Authors: | , , |
Format: | Journal Article |
Language: | English |
Published: |
New York, NY
IEEE
01-03-1992
Institute of Electrical and Electronics Engineers |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A heart-sounds gating device has been designed and tested which identifies, individually, both the first (S1) and second (S2) heart sound from their timing relationship, providing two trigger points through the cardiac cycle for synchronizing medical images. The gate utilizes dynamically varying timing windows to anticipate the occurrence of S1 and S2. It has been initially applied to nuclear imaging of the cardiac bloodpool, but may be applied to any imaging modality requiring cardiac synchronization.< > |
---|---|
Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 ObjectType-Article-1 ObjectType-Feature-2 |
ISSN: | 0018-9294 1558-2531 |
DOI: | 10.1109/10.125016 |