Measurement of beam-axis displacement from the isocenter during three-dimensional conformal radiosurgery with a micro-multileaf collimator

We describe the displacement of the beam-axis from the planning isocenter in clinical situations during three-dimensional conformal radiosurgery using an Acculeaf bi-directional micro-multileaf collimator. The displacements were recorded for 64 ports using a video imaging system and a stereotactic a...

Full description

Saved in:
Bibliographic Details
Published in:Radiotherapy and oncology Vol. 70; no. 1; pp. 45 - 48
Main Authors: Kunieda, Etsuo, Kawaguchi, Osamu, Saitoh, Hidetoshi, Fujisaki, Tatsuya, Takeda, Atsuya, Kawase, Takatsugu, Deloar, Hossain M., Shigematsu, Naoyuki, Kubo, Atsushi
Format: Journal Article
Language:English
Published: Ireland Elsevier Ireland Ltd 2004
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We describe the displacement of the beam-axis from the planning isocenter in clinical situations during three-dimensional conformal radiosurgery using an Acculeaf bi-directional micro-multileaf collimator. The displacements were recorded for 64 ports using a video imaging system and a stereotactic arc. The mean displacement was 0.41±0.25 mm.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0167-8140
1879-0887
DOI:10.1016/j.radonc.2003.11.008