Management of fractured dental implants: a case report
An implant fracture may be one of the major causes of late implant failures. Complications, such as loosening or fracture of the prosthesis restorative components, or even fracture of the implant, may occur and dental professionals should be aware of the causes of these complications. This study rep...
Saved in:
Published in: | Implant dentistry Vol. 18; no. 1; pp. 10 - 16 |
---|---|
Main Authors: | , , , |
Format: | Journal Article |
Language: | English |
Published: |
United States
01-02-2009
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An implant fracture may be one of the major causes of late implant failures. Complications, such as loosening or fracture of the prosthesis restorative components, or even fracture of the implant, may occur and dental professionals should be aware of the causes of these complications. This study reports a clinical situation involving a patient restored with a mandibular overdenture that presented a fractured implant 2 years after placement. The probable cause of the implant fracture was due to biomechanical overload caused by parafunctional habits. The implant head was flattened to make it smooth, retapping the internal screw, installing a new abutment (longer), and fabricating part of the overdenture bar. This treatment was timesaving for the patient in that the prosthesis was repaired in the shortest time possible. |
---|---|
Bibliography: | ObjectType-Case Study-2 SourceType-Scholarly Journals-1 ObjectType-Feature-4 content type line 23 ObjectType-Report-1 ObjectType-Article-3 |
ISSN: | 1056-6163 1538-2982 |
DOI: | 10.1097/ID.0b013e318192cafe |