Artefactual 25-OH vitamin D concentration in multiple myeloma
The most commonly used techniques to measure vitamin D are automated immunoassays which are known to be affected by interferences, especially from immunoglobulins present in the patient's serum. We present a case of a patient with myeloma in whom interference with the vitamin D assay was identi...
Saved in:
Published in: | Annals of clinical biochemistry Vol. 54; no. 6; p. 716 |
---|---|
Main Authors: | , , , , |
Format: | Journal Article |
Language: | English |
Published: |
England
01-11-2017
|
Subjects: | |
Online Access: | Get more information |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The most commonly used techniques to measure vitamin D are automated immunoassays which are known to be affected by interferences, especially from immunoglobulins present in the patient's serum. We present a case of a patient with myeloma in whom interference with the vitamin D assay was identified. An 83-year-old female, known to have IgG myeloma, was found to have a high concentration of 25-OH vitamin D on a routine test without any signs of vitamin D toxicity. She was not taking vitamin D supplements or any other multivitamin preparation and had minimal sun exposure. The initial and subsequent samples run by the ARCHITECT 25-OH vitamin D assay (chemiluminescent microparticle immunoassay technology, Abbott Laboratories, Abbott Park, IL) showed a high concentration of 25-OH vitamin D of 281 nmol/L and 327 nmol/L, respectively. Further fresh samples taken for 25-OH vitamin D and analysed by liquid chromatography-mass spectrometry (LC-MS/MS) and ARCHITECT analysis showed results of 49 nmol/L and 289 nmol/L, respectively. Our patient had high concentrations of circulating IgG paraproteins and had a long history of rheumatoid arthritis; paraproteins and rheumatoid factor may interfere in the assay. In conclusion, we report a case of a patient with IgG myeloma and rheumatoid arthritis with high concentrations of 25-OH vitamin D detected by the Abbott ARCHITECT, but not by a reference method (LC-MS/MS). The most likely cause of the discordant results is interference in the immunoassay by the paraprotein but interference from rheumatoid factor remains a possibility. |
---|---|
ISSN: | 1758-1001 |
DOI: | 10.1177/0004563217690175 |