Calibration procedure of the ΔE–E detectors for dp breakup investigation at nuclotron

Calibration procedure of Δ E − E detectors used in dp breakup reaction measurement in the framework of DSS project is discussed. Time information from all PMTs along with amplitude information and known energy values of pp quasi elastic reaction are used to find calibration coefficients for Δ E and...

Full description

Saved in:
Bibliographic Details
Published in:Physics of particles and nuclei letters Vol. 15; no. 1; pp. 76 - 82
Main Authors: Janek, M., Ladygin, V. P., Piyadin, S. M., Gurchin, Yu. V., Isupov, A. Yu, Karachuk, J. T., Khrenov, A. N., Kurilkin, A. K., Kurilkin, P. K., Livanov, A. N., Martinska, G., Reznikov, S. G., Tarjanyiova, G., Terekhin, A. A.
Format: Journal Article
Language:English
Published: Moscow Pleiades Publishing 2018
Springer Nature B.V
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Calibration procedure of Δ E − E detectors used in dp breakup reaction measurement in the framework of DSS project is discussed. Time information from all PMTs along with amplitude information and known energy values of pp quasi elastic reaction are used to find calibration coefficients for Δ E and E detectors. Calibration coefficients are used to recover deposited particle energies. Δ E vs. E plots, energies and missing mass spectra are compared with geant4 Monte Carlo simulation. Missing mass for particular physical configuration is calculated as a test of the calibration procedure quality.
ISSN:1547-4771
1531-8567
DOI:10.1134/S1547477118010090