Determination of alpha-tocopherol in rat liver by high-performance liquid chromatography utilizing ultraviolet detection
A rapid, sensitive analytical procedure was developed to quantitate alpha-tocopherol in rat liver. Liver is homogenized in acetone, fractionated by reversed-phase high-performance liquid chromatography and alpha-tocopherol quantitated by monitoring the column effluent at 280 nm. A single analysis re...
Saved in:
Published in: | Journal of chromatography Vol. 344; p. 137 |
---|---|
Main Authors: | , , , |
Format: | Journal Article |
Language: | English |
Published: |
Netherlands
01-01-1985
|
Subjects: | |
Online Access: | Get more information |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A rapid, sensitive analytical procedure was developed to quantitate alpha-tocopherol in rat liver. Liver is homogenized in acetone, fractionated by reversed-phase high-performance liquid chromatography and alpha-tocopherol quantitated by monitoring the column effluent at 280 nm. A single analysis requires 18 min. The method is linear from 1.0 to greater than 30.0 micrograms alpha-tocopherol per g of liver (wet weight). The average relative standard deviation of the slope of the standard curve is 2.3% and the single-day coefficient of variation is 6.3%. The use of an acetone extraction, reversed-phase column chromatography and quantitation of alpha-tocopherol at 280 nm results in a sensitive and reproducible system for the determination of alpha-tocopherol concentrations in rat liver. Preliminary studies have shown the assay to be useful for investigation of the effects of age and diet on hepatic alpha-tocopherol concentration. |
---|---|
DOI: | 10.1016/S0378-4347(00)82014-9 |