A Study of 3D-NPS Analysis in CT Images Based on the Central Cross-section Theorem
The noise power spectrum (NPS) of a CT scanner is commonly measured from a single noise image. However, since CT images are three-dimensional (3D) volume data, they have 3D noise characteristics (3D-NPS). In this study, we clarify the relationship among NPSs measured by various approaches in NPS ana...
Saved in:
Published in: | Nihon Hoshasen Gijutsu Gakkai zasshi Vol. 78; no. 4; p. 342 |
---|---|
Main Authors: | , , , , |
Format: | Journal Article |
Language: | English Japanese |
Published: |
Japan
01-01-2022
|
Subjects: | |
Online Access: | Get more information |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Abstract | The noise power spectrum (NPS) of a CT scanner is commonly measured from a single noise image. However, since CT images are three-dimensional (3D) volume data, they have 3D noise characteristics (3D-NPS). In this study, we clarify the relationship among NPSs measured by various approaches in NPS analysis based on the central slice theorem. Its validity is verified by the NPS measurements using actual 3D noise data.
We defined the NPS
(f
, f
) that was calculated by the 2D Fourier transform (FT) from the 2D projection of 3D noise data in the patient longitudinal direction, the 3D-NPS(f
, f
, f
) that was calculated by the 3D-FT from the 3D noise data, and the 2D-NPS(f
, f
) that was calculated by the 2D-FT from a single noise image; f
, f
, and f
are spatial frequencies corresponding to the axes of x, y, and z in the reconstructed CT volume, respectively. Based on the central slice theorem, we described that the NPS
(f
, f
=0) was equal to the 3D-NPS(f
, f
=0, f
=0), and the NPS(2D-NPS(f
, f
=0)) was different from the 3D-NPS(f
, f
=0, f
=0). To verify them, we compared the NPSs calculated from actual 3D noise data that were obtained using a cylindrical water phantom.
The 3D-NPS(f
, f
=0, f
=0) matched the NPS
(f
, f
=0) and was different from the 2D-NPS(f
, f
=0).
Based on the central slice theorem, we clarified the relationship among NPSs measured by various approaches in NPS analysis; it is important to understand this and then select an appropriate noise data handling and NPS measurement method. |
---|---|
AbstractList | The noise power spectrum (NPS) of a CT scanner is commonly measured from a single noise image. However, since CT images are three-dimensional (3D) volume data, they have 3D noise characteristics (3D-NPS). In this study, we clarify the relationship among NPSs measured by various approaches in NPS analysis based on the central slice theorem. Its validity is verified by the NPS measurements using actual 3D noise data.
We defined the NPS
(f
, f
) that was calculated by the 2D Fourier transform (FT) from the 2D projection of 3D noise data in the patient longitudinal direction, the 3D-NPS(f
, f
, f
) that was calculated by the 3D-FT from the 3D noise data, and the 2D-NPS(f
, f
) that was calculated by the 2D-FT from a single noise image; f
, f
, and f
are spatial frequencies corresponding to the axes of x, y, and z in the reconstructed CT volume, respectively. Based on the central slice theorem, we described that the NPS
(f
, f
=0) was equal to the 3D-NPS(f
, f
=0, f
=0), and the NPS(2D-NPS(f
, f
=0)) was different from the 3D-NPS(f
, f
=0, f
=0). To verify them, we compared the NPSs calculated from actual 3D noise data that were obtained using a cylindrical water phantom.
The 3D-NPS(f
, f
=0, f
=0) matched the NPS
(f
, f
=0) and was different from the 2D-NPS(f
, f
=0).
Based on the central slice theorem, we clarified the relationship among NPSs measured by various approaches in NPS analysis; it is important to understand this and then select an appropriate noise data handling and NPS measurement method. |
Author | Noto, Yoshiyuki Narita, Akihiro Ohkubo, Masaki Fukaya, Takahiro Sakai, Kenichi |
Author_xml | – sequence: 1 givenname: Akihiro surname: Narita fullname: Narita, Akihiro organization: Department of Radiological Technology, Graduate School of Health Sciences, Niigata University – sequence: 2 givenname: Masaki surname: Ohkubo fullname: Ohkubo, Masaki organization: Department of Radiological Technology, Graduate School of Health Sciences, Niigata University – sequence: 3 givenname: Takahiro surname: Fukaya fullname: Fukaya, Takahiro organization: Division of Radiology, Niigata University Medical and Dental Hospital – sequence: 4 givenname: Kenichi surname: Sakai fullname: Sakai, Kenichi organization: Division of Radiology, Niigata University Medical and Dental Hospital – sequence: 5 givenname: Yoshiyuki surname: Noto fullname: Noto, Yoshiyuki organization: Division of Radiology, Niigata University Medical and Dental Hospital |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/35228483$$D View this record in MEDLINE/PubMed |
BookMark | eNo1j8tKw0AUQAdRbK3dupT5gakzN_NcxvgqFBUb12WSubEpeZRMusjfK6irszhw4FyR867vkJAbwVeac3d3OAzjCjgAEyDMGZkLawWT1iYzsoyxLjgHaaVS6pLMEgVgpU3m5COl2_EUJtpXNHlgr-9bmna-mWIdad3RLKfr1n9hpPc-YqB9R8c90gy7cfANzYY-RhaxHOsfk--xH7C9JheVbyIu_7ggn0-PefbCNm_P6yzdsBJAGuaC1y6gUVUwpRFFGRyIylu0WGonlEs8rxzICrk06LUHE5QGz1Gg0lrDgtz-do-nosWwOw5164dp9z8H3wHHUHo |
ContentType | Journal Article |
DBID | CGR CUY CVF ECM EIF NPM |
DOI | 10.6009/jjrt.2022-1217 |
DatabaseName | Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed |
DatabaseTitle | MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) |
DatabaseTitleList | MEDLINE |
Database_xml | – sequence: 1 dbid: ECM name: MEDLINE url: https://search.ebscohost.com/login.aspx?direct=true&db=cmedm&site=ehost-live sourceTypes: Index Database |
DeliveryMethod | no_fulltext_linktorsrc |
EISSN | 1881-4883 |
ExternalDocumentID | 35228483 |
Genre | Journal Article |
GroupedDBID | CGR CUY CVF ECM EIF NPM |
ID | FETCH-LOGICAL-c2247-9da69de75fd7c71bcd921fa8e8ec691593a0f924fe047ea6a27d562a0e1e56662 |
IngestDate | Thu Nov 24 22:03:11 EST 2022 |
IsDoiOpenAccess | false |
IsOpenAccess | true |
IsPeerReviewed | false |
IsScholarly | true |
Issue | 4 |
Keywords | central slice theorem image noise projection slice theorem computed tomography (CT) noise power spectrum (NPS) |
Language | English Japanese |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c2247-9da69de75fd7c71bcd921fa8e8ec691593a0f924fe047ea6a27d562a0e1e56662 |
OpenAccessLink | https://www.jstage.jst.go.jp/article/jjrt/78/4/78_2022-1217/_article/-char/en |
PMID | 35228483 |
ParticipantIDs | pubmed_primary_35228483 |
PublicationCentury | 2000 |
PublicationDate | 20220101 |
PublicationDateYYYYMMDD | 2022-01-01 |
PublicationDate_xml | – month: 01 year: 2022 text: 20220101 day: 01 |
PublicationDecade | 2020 |
PublicationPlace | Japan |
PublicationPlace_xml | – name: Japan |
PublicationTitle | Nihon Hoshasen Gijutsu Gakkai zasshi |
PublicationTitleAlternate | Nihon Hoshasen Gijutsu Gakkai Zasshi |
PublicationYear | 2022 |
SSID | ssib002484555 |
Score | 2.248214 |
Snippet | The noise power spectrum (NPS) of a CT scanner is commonly measured from a single noise image. However, since CT images are three-dimensional (3D) volume data,... |
SourceID | pubmed |
SourceType | Index Database |
StartPage | 342 |
SubjectTerms | Algorithms Humans Image Processing, Computer-Assisted - methods Phantoms, Imaging Signal-To-Noise Ratio Tomography Scanners, X-Ray Computed Tomography, X-Ray Computed - methods |
Title | A Study of 3D-NPS Analysis in CT Images Based on the Central Cross-section Theorem |
URI | https://www.ncbi.nlm.nih.gov/pubmed/35228483 |
Volume | 78 |
hasFullText | |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://sdu.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lj9MwELZakNBeEIg3u8gHbpFF3naO3W53y4EK0SBxWzmxo7jRJqsNOcCvZ_xIm60EggOXKLHlxMp8GX8z8cwg9D5ilAVRVZCAsoLEsISRIi0yUqYZkzyGFc9k119v6eYbu1jFq9lsLLN1aPuvkoY2kLWOnP0Hae9vCg1wDjKHI0gdjn8l94XZGmj-m0cXZPN5e8g7ogP8cu_jDdd5Hc5h-RLuV8Ho4_WWes0kvbT1w03cvnPtOv66UTV0rLu-huGtd6V2w_d-8K5403Dl_QQmXqu9f5lrx4NRPo2q1V23d-jWzVB0NlKoBwa7x9DQ8B9mQM4bPh2xhWvlwohU6R7hfBVhOPFVSKtfGQOTldnaNaMCpmwCtHiiTSObeOtYywNH00lSd7s7vRk21MVZbPznRLy3N0a-ml6y2D7uz71HWbfHrjmaA4fSNHv5aUJrWJwkic0Bqmfz4f5cTtCjcfyRtWJYS_4EPXbmBl5YnDxFsx1_hr4ssMEI7ipsMYJHjGDV4mWOLUawwQjuWgwYwQ4j-B5GsMPIc_T1cpUv18TV1iAlkDZKMsHTTEiaVIKWNChKkYVBxZlkEj5T4LgR9yuwzSvpx1TylIdUAFXmvgwkWABp-AI9aLtWvkI49QUPRQJ0iBWa7xUR47zwQ8EEkJ04e41e2ldwfWsTqFyPL-fNb3veopMDfk7Rwwq-TnmG5r0Y3hlZ_AIifFeg |
link.rule.ids | 782 |
linkProvider | EBSCOhost |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=A+Study+of+3D-NPS+Analysis+in+CT+Images+Based+on+the+Central+Cross-section+Theorem&rft.jtitle=Nihon+Hoshasen+Gijutsu+Gakkai+zasshi&rft.au=Narita%2C+Akihiro&rft.au=Ohkubo%2C+Masaki&rft.au=Fukaya%2C+Takahiro&rft.au=Sakai%2C+Kenichi&rft.date=2022-01-01&rft.eissn=1881-4883&rft.volume=78&rft.issue=4&rft.spage=342&rft_id=info:doi/10.6009%2Fjjrt.2022-1217&rft_id=info%3Apmid%2F35228483&rft_id=info%3Apmid%2F35228483&rft.externalDocID=35228483 |