A Study of 3D-NPS Analysis in CT Images Based on the Central Cross-section Theorem

The noise power spectrum (NPS) of a CT scanner is commonly measured from a single noise image. However, since CT images are three-dimensional (3D) volume data, they have 3D noise characteristics (3D-NPS). In this study, we clarify the relationship among NPSs measured by various approaches in NPS ana...

Full description

Saved in:
Bibliographic Details
Published in:Nihon Hoshasen Gijutsu Gakkai zasshi Vol. 78; no. 4; p. 342
Main Authors: Narita, Akihiro, Ohkubo, Masaki, Fukaya, Takahiro, Sakai, Kenichi, Noto, Yoshiyuki
Format: Journal Article
Language:English
Japanese
Published: Japan 01-01-2022
Subjects:
Online Access:Get more information
Tags: Add Tag
No Tags, Be the first to tag this record!
Abstract The noise power spectrum (NPS) of a CT scanner is commonly measured from a single noise image. However, since CT images are three-dimensional (3D) volume data, they have 3D noise characteristics (3D-NPS). In this study, we clarify the relationship among NPSs measured by various approaches in NPS analysis based on the central slice theorem. Its validity is verified by the NPS measurements using actual 3D noise data. We defined the NPS (f , f ) that was calculated by the 2D Fourier transform (FT) from the 2D projection of 3D noise data in the patient longitudinal direction, the 3D-NPS(f , f , f ) that was calculated by the 3D-FT from the 3D noise data, and the 2D-NPS(f , f ) that was calculated by the 2D-FT from a single noise image; f , f , and f are spatial frequencies corresponding to the axes of x, y, and z in the reconstructed CT volume, respectively. Based on the central slice theorem, we described that the NPS (f , f =0) was equal to the 3D-NPS(f , f =0, f =0), and the NPS(2D-NPS(f , f =0)) was different from the 3D-NPS(f , f =0, f =0). To verify them, we compared the NPSs calculated from actual 3D noise data that were obtained using a cylindrical water phantom. The 3D-NPS(f , f =0, f =0) matched the NPS (f , f =0) and was different from the 2D-NPS(f , f =0). Based on the central slice theorem, we clarified the relationship among NPSs measured by various approaches in NPS analysis; it is important to understand this and then select an appropriate noise data handling and NPS measurement method.
AbstractList The noise power spectrum (NPS) of a CT scanner is commonly measured from a single noise image. However, since CT images are three-dimensional (3D) volume data, they have 3D noise characteristics (3D-NPS). In this study, we clarify the relationship among NPSs measured by various approaches in NPS analysis based on the central slice theorem. Its validity is verified by the NPS measurements using actual 3D noise data. We defined the NPS (f , f ) that was calculated by the 2D Fourier transform (FT) from the 2D projection of 3D noise data in the patient longitudinal direction, the 3D-NPS(f , f , f ) that was calculated by the 3D-FT from the 3D noise data, and the 2D-NPS(f , f ) that was calculated by the 2D-FT from a single noise image; f , f , and f are spatial frequencies corresponding to the axes of x, y, and z in the reconstructed CT volume, respectively. Based on the central slice theorem, we described that the NPS (f , f =0) was equal to the 3D-NPS(f , f =0, f =0), and the NPS(2D-NPS(f , f =0)) was different from the 3D-NPS(f , f =0, f =0). To verify them, we compared the NPSs calculated from actual 3D noise data that were obtained using a cylindrical water phantom. The 3D-NPS(f , f =0, f =0) matched the NPS (f , f =0) and was different from the 2D-NPS(f , f =0). Based on the central slice theorem, we clarified the relationship among NPSs measured by various approaches in NPS analysis; it is important to understand this and then select an appropriate noise data handling and NPS measurement method.
Author Noto, Yoshiyuki
Narita, Akihiro
Ohkubo, Masaki
Fukaya, Takahiro
Sakai, Kenichi
Author_xml – sequence: 1
  givenname: Akihiro
  surname: Narita
  fullname: Narita, Akihiro
  organization: Department of Radiological Technology, Graduate School of Health Sciences, Niigata University
– sequence: 2
  givenname: Masaki
  surname: Ohkubo
  fullname: Ohkubo, Masaki
  organization: Department of Radiological Technology, Graduate School of Health Sciences, Niigata University
– sequence: 3
  givenname: Takahiro
  surname: Fukaya
  fullname: Fukaya, Takahiro
  organization: Division of Radiology, Niigata University Medical and Dental Hospital
– sequence: 4
  givenname: Kenichi
  surname: Sakai
  fullname: Sakai, Kenichi
  organization: Division of Radiology, Niigata University Medical and Dental Hospital
– sequence: 5
  givenname: Yoshiyuki
  surname: Noto
  fullname: Noto, Yoshiyuki
  organization: Division of Radiology, Niigata University Medical and Dental Hospital
BackLink https://www.ncbi.nlm.nih.gov/pubmed/35228483$$D View this record in MEDLINE/PubMed
BookMark eNo1j8tKw0AUQAdRbK3dupT5gakzN_NcxvgqFBUb12WSubEpeZRMusjfK6irszhw4FyR867vkJAbwVeac3d3OAzjCjgAEyDMGZkLawWT1iYzsoyxLjgHaaVS6pLMEgVgpU3m5COl2_EUJtpXNHlgr-9bmna-mWIdad3RLKfr1n9hpPc-YqB9R8c90gy7cfANzYY-RhaxHOsfk--xH7C9JheVbyIu_7ggn0-PefbCNm_P6yzdsBJAGuaC1y6gUVUwpRFFGRyIylu0WGonlEs8rxzICrk06LUHE5QGz1Gg0lrDgtz-do-nosWwOw5164dp9z8H3wHHUHo
ContentType Journal Article
DBID CGR
CUY
CVF
ECM
EIF
NPM
DOI 10.6009/jjrt.2022-1217
DatabaseName Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
DatabaseTitle MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
DatabaseTitleList MEDLINE
Database_xml – sequence: 1
  dbid: ECM
  name: MEDLINE
  url: https://search.ebscohost.com/login.aspx?direct=true&db=cmedm&site=ehost-live
  sourceTypes: Index Database
DeliveryMethod no_fulltext_linktorsrc
EISSN 1881-4883
ExternalDocumentID 35228483
Genre Journal Article
GroupedDBID CGR
CUY
CVF
ECM
EIF
NPM
ID FETCH-LOGICAL-c2247-9da69de75fd7c71bcd921fa8e8ec691593a0f924fe047ea6a27d562a0e1e56662
IngestDate Thu Nov 24 22:03:11 EST 2022
IsDoiOpenAccess false
IsOpenAccess true
IsPeerReviewed false
IsScholarly true
Issue 4
Keywords central slice theorem
image noise
projection slice theorem
computed tomography (CT)
noise power spectrum (NPS)
Language English
Japanese
LinkModel OpenURL
MergedId FETCHMERGED-LOGICAL-c2247-9da69de75fd7c71bcd921fa8e8ec691593a0f924fe047ea6a27d562a0e1e56662
OpenAccessLink https://www.jstage.jst.go.jp/article/jjrt/78/4/78_2022-1217/_article/-char/en
PMID 35228483
ParticipantIDs pubmed_primary_35228483
PublicationCentury 2000
PublicationDate 20220101
PublicationDateYYYYMMDD 2022-01-01
PublicationDate_xml – month: 01
  year: 2022
  text: 20220101
  day: 01
PublicationDecade 2020
PublicationPlace Japan
PublicationPlace_xml – name: Japan
PublicationTitle Nihon Hoshasen Gijutsu Gakkai zasshi
PublicationTitleAlternate Nihon Hoshasen Gijutsu Gakkai Zasshi
PublicationYear 2022
SSID ssib002484555
Score 2.248214
Snippet The noise power spectrum (NPS) of a CT scanner is commonly measured from a single noise image. However, since CT images are three-dimensional (3D) volume data,...
SourceID pubmed
SourceType Index Database
StartPage 342
SubjectTerms Algorithms
Humans
Image Processing, Computer-Assisted - methods
Phantoms, Imaging
Signal-To-Noise Ratio
Tomography Scanners, X-Ray Computed
Tomography, X-Ray Computed - methods
Title A Study of 3D-NPS Analysis in CT Images Based on the Central Cross-section Theorem
URI https://www.ncbi.nlm.nih.gov/pubmed/35228483
Volume 78
hasFullText
inHoldings 1
isFullTextHit
isPrint
link http://sdu.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lj9MwELZakNBeEIg3u8gHbpFF3naO3W53y4EK0SBxWzmxo7jRJqsNOcCvZ_xIm60EggOXKLHlxMp8GX8z8cwg9D5ilAVRVZCAsoLEsISRIi0yUqYZkzyGFc9k119v6eYbu1jFq9lsLLN1aPuvkoY2kLWOnP0Hae9vCg1wDjKHI0gdjn8l94XZGmj-m0cXZPN5e8g7ogP8cu_jDdd5Hc5h-RLuV8Ho4_WWes0kvbT1w03cvnPtOv66UTV0rLu-huGtd6V2w_d-8K5403Dl_QQmXqu9f5lrx4NRPo2q1V23d-jWzVB0NlKoBwa7x9DQ8B9mQM4bPh2xhWvlwohU6R7hfBVhOPFVSKtfGQOTldnaNaMCpmwCtHiiTSObeOtYywNH00lSd7s7vRk21MVZbPznRLy3N0a-ml6y2D7uz71HWbfHrjmaA4fSNHv5aUJrWJwkic0Bqmfz4f5cTtCjcfyRtWJYS_4EPXbmBl5YnDxFsx1_hr4ssMEI7ipsMYJHjGDV4mWOLUawwQjuWgwYwQ4j-B5GsMPIc_T1cpUv18TV1iAlkDZKMsHTTEiaVIKWNChKkYVBxZlkEj5T4LgR9yuwzSvpx1TylIdUAFXmvgwkWABp-AI9aLtWvkI49QUPRQJ0iBWa7xUR47zwQ8EEkJ04e41e2ldwfWsTqFyPL-fNb3veopMDfk7Rwwq-TnmG5r0Y3hlZ_AIifFeg
link.rule.ids 782
linkProvider EBSCOhost
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=A+Study+of+3D-NPS+Analysis+in+CT+Images+Based+on+the+Central+Cross-section+Theorem&rft.jtitle=Nihon+Hoshasen+Gijutsu+Gakkai+zasshi&rft.au=Narita%2C+Akihiro&rft.au=Ohkubo%2C+Masaki&rft.au=Fukaya%2C+Takahiro&rft.au=Sakai%2C+Kenichi&rft.date=2022-01-01&rft.eissn=1881-4883&rft.volume=78&rft.issue=4&rft.spage=342&rft_id=info:doi/10.6009%2Fjjrt.2022-1217&rft_id=info%3Apmid%2F35228483&rft_id=info%3Apmid%2F35228483&rft.externalDocID=35228483