Inconsistencies of neutron flux parameters for k0 standardization in neutron activation analysis determined with the use of Au+Zr and Au+Mo+Cr monitor sets at the LVR-15 reactor in Řež

Multipurpose research reactors require monitoring of neutron flux parameters ( α , f , F c or Φ th ) during every irradiation of samples when k 0 standardization in neutron activation analysis is to be used, because the above parameters may change quite frequently and unpredictably. We tested two mo...

Full description

Saved in:
Bibliographic Details
Published in:Journal of radioanalytical and nuclear chemistry Vol. 293; no. 2; pp. 665 - 674
Main Authors: Kubešová, Marie, Kučera, Jan
Format: Journal Article
Language:English
Published: Dordrecht Springer Netherlands 01-08-2012
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Multipurpose research reactors require monitoring of neutron flux parameters ( α , f , F c or Φ th ) during every irradiation of samples when k 0 standardization in neutron activation analysis is to be used, because the above parameters may change quite frequently and unpredictably. We tested two monitor sets consisting of Au+Zr and Au+Mo+Cr for determination of the neutron flux parameters in k 0 -NAA at the LVR-15 reactor in Řež. Based on analysis of several reference materials, we found the use of the Au+Zr monitor set superior over the Au+Mo+Cr monitor set in the specific conditions of the LVR-15 reactor.
ISSN:0236-5731
1588-2780
DOI:10.1007/s10967-012-1709-y