SU‐GG‐J‐175: Target Registration Error Analysis Via KV Imaging and Conebeam CT in Accelerated Partial Breast Irradiation
Purpose: In accelerated partial breast irradiation (APBI), accurate localization of the lumpectomy cavity is important. We are investigating daily localization of the lumpectomy cavity in patients with implanted surgical clips by kV imaging followed by conebeam CT (CBCT). Initial data based on 2 pat...
Saved in:
Published in: | Medical Physics Vol. 35; no. 6; p. 2720 |
---|---|
Main Authors: | , , , , , , |
Format: | Conference Proceeding Journal Article |
Language: | English |
Published: |
American Association of Physicists in Medicine
01-06-2008
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Abstract | Purpose: In accelerated partial breast irradiation (APBI), accurate localization of the lumpectomy cavity is important. We are investigating daily localization of the lumpectomy cavity in patients with implanted surgical clips by kV imaging followed by conebeam CT (CBCT). Initial data based on 2 patients are presented here. Target registration error (TRE) analysis was performed to assess the accuracy of available image‐guidance techniques for target localization during APBI. Method and Materials: APBI patients were positioned to the treatment isocenter initially by laser alignment to skin tattoos, followed by orthogonal kV radiographs to visualize bony anatomy and surgical clips and then by CBCT (once daily) to visualize both surgical clips and soft tissue architecture. Couch shifts were performed based on matching surgical clip position and lumpectomy cavity architecture (when visible) to the planned setup. The lumpectomy cavity position was assumed to be represented by the surgical clips. The TRE is defined as the residual setup error following isocenter alignment. Surface topology analysis on the serial CBCT datasets was performed post‐treatment by co‐registering the breast surfaces, as defined by CBCT, and then measuring the clip misalignment. Results: The average laser setup TREs (determined by kV imaging) for patients 1 (breast volume = 1335.5cc) and 2 (breast volume = 404.7cc) were 3.7mm and 6.1mm, respectively. The average kV imaging TREs (determined by CBCT, post‐kV imaging shifts) for patients 1 and 2 were 4.9mm and 1.4mm, respectively. The average surface topology TREs for patients 1 and 2 were 4.9mm and 1.6mm, respectively. Conclusion: Even after kV imaging, TREs around 0.5cm were detected using CBCT in the larger breast volume. Therefore, CBCT may improve targeting in APBI for patients with large breast volumes. Surface alignment may be more suitable for patients with small breast volumes and is currently under further investigation. |
---|---|
AbstractList | Purpose: In accelerated partial breast irradiation (APBI), accurate localization of the lumpectomy cavity is important. We are investigating daily localization of the lumpectomy cavity in patients with implanted surgical clips by kV imaging followed by conebeam CT (CBCT). Initial data based on 2 patients are presented here. Target registration error (TRE) analysis was performed to assess the accuracy of available image‐guidance techniques for target localization during APBI. Method and Materials: APBI patients were positioned to the treatment isocenter initially by laser alignment to skin tattoos, followed by orthogonal kV radiographs to visualize bony anatomy and surgical clips and then by CBCT (once daily) to visualize both surgical clips and soft tissue architecture. Couch shifts were performed based on matching surgical clip position and lumpectomy cavity architecture (when visible) to the planned setup. The lumpectomy cavity position was assumed to be represented by the surgical clips. The TRE is defined as the residual setup error following isocenter alignment. Surface topology analysis on the serial CBCT datasets was performed post‐treatment by co‐registering the breast surfaces, as defined by CBCT, and then measuring the clip misalignment. Results: The average laser setup TREs (determined by kV imaging) for patients 1 (breast volume = 1335.5cc) and 2 (breast volume = 404.7cc) were 3.7mm and 6.1mm, respectively. The average kV imaging TREs (determined by CBCT, post‐kV imaging shifts) for patients 1 and 2 were 4.9mm and 1.4mm, respectively. The average surface topology TREs for patients 1 and 2 were 4.9mm and 1.6mm, respectively. Conclusion: Even after kV imaging, TREs around 0.5cm were detected using CBCT in the larger breast volume. Therefore, CBCT may improve targeting in APBI for patients with large breast volumes. Surface alignment may be more suitable for patients with small breast volumes and is currently under further investigation. Abstract only Purpose: In accelerated partial breast irradiation (APBI), accurate localization of the lumpectomy cavity is important. We are investigating daily localization of the lumpectomy cavity in patients with implanted surgical clips by kV imaging followed by conebeam CT (CBCT). Initial data based on 2 patients are presented here. Target registration error (TRE) analysis was performed to assess the accuracy of available image‐guidance techniques for target localization during APBI. Method and Materials: APBI patients were positioned to the treatment isocenter initially by laser alignment to skin tattoos, followed by orthogonal kV radiographs to visualize bony anatomy and surgical clips and then by CBCT (once daily) to visualize both surgical clips and soft tissue architecture. Couch shifts were performed based on matching surgical clip position and lumpectomy cavity architecture (when visible) to the planned setup. The lumpectomy cavity position was assumed to be represented by the surgical clips. The TRE is defined as the residual setup error following isocenter alignment. Surface topology analysis on the serial CBCT datasets was performed post‐treatment by co‐registering the breast surfaces, as defined by CBCT, and then measuring the clip misalignment. Results: The average laser setup TREs (determined by kV imaging) for patients 1 (breast volume = 1335.5cc) and 2 (breast volume = 404.7cc) were 3.7mm and 6.1mm, respectively. The average kV imaging TREs (determined by CBCT, post‐kV imaging shifts) for patients 1 and 2 were 4.9mm and 1.4mm, respectively. The average surface topology TREs for patients 1 and 2 were 4.9mm and 1.6mm, respectively. Conclusion: Even after kV imaging, TREs around 0.5cm were detected using CBCT in the larger breast volume. Therefore, CBCT may improve targeting in APBI for patients with large breast volumes. Surface alignment may be more suitable for patients with small breast volumes and is currently under further investigation. |
Author | Kachnic, L Schuller, B Nawaz, A Gierga, D Wolfgang, J Willins, J Chen, G |
Author_xml | – sequence: 1 givenname: B surname: Schuller fullname: Schuller, B organization: 2Massachusetts General Hospital, Boston, MA – sequence: 2 givenname: A surname: Nawaz fullname: Nawaz, A organization: 2Massachusetts General Hospital, Boston, MA – sequence: 3 givenname: J surname: Wolfgang fullname: Wolfgang, J organization: 2Massachusetts General Hospital, Boston, MA – sequence: 4 givenname: D surname: Gierga fullname: Gierga, D organization: 2Massachusetts General Hospital, Boston, MA – sequence: 5 givenname: J surname: Willins fullname: Willins, J organization: 2Massachusetts General Hospital, Boston, MA – sequence: 6 givenname: L surname: Kachnic fullname: Kachnic, L organization: 2Massachusetts General Hospital, Boston, MA – sequence: 7 givenname: G surname: Chen fullname: Chen, G organization: 2Massachusetts General Hospital, Boston, MA |
BookMark | eNp9kLFOwzAQhi1UJNrCwBt4BSnFdhwnZitRKYUiKmi7Ro5zjozSBNmRUBfEI_CMPAmh7QrD3S3f_br7BqhXNzUgdE7JiFKaXNERk4LGjB-hPuNxGHBGZA_1CZE8YJxEJ2jg_SshRIQR6aOPl9X359d02rX7rmgcXeOlciW0-BlK61unWtvUeOJc4_C4VtXWW4_XVuGHNZ5tVGnrEqu6wGl3SA5qg9MltjUeaw0VdNtQ4IVyrVUVvnGgfItnzqnC7nJP0bFRlYezwxyi1e1kmd4F86fpLB3PA03jmAeguEgMZbnmBiISgQh5mEDOSRxKJbURSWhkTLgpiDGdAc0ianIhJBNJbiAcoot9rnaN9w5M9ubsRrltRkn2Ky6j2UFcxwZ79t1WsP0bzB4XB_5yz3tt291X_4T_AEKlfrg |
CODEN | MPHYA6 |
ContentType | Conference Proceeding Journal Article |
Copyright | American Association of Physicists in Medicine 2008 American Association of Physicists in Medicine |
Copyright_xml | – notice: American Association of Physicists in Medicine – notice: 2008 American Association of Physicists in Medicine |
DBID | AAYXX CITATION |
DOI | 10.1118/1.2961724 |
DatabaseName | CrossRef |
DatabaseTitle | CrossRef |
DatabaseTitleList | CrossRef |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Medicine Physics |
EISSN | 2473-4209 |
EndPage | 2720 |
ExternalDocumentID | 10_1118_1_2961724 MP1724 |
Genre | abstract |
GroupedDBID | --- --Z -DZ .GJ 0R~ 1OB 1OC 29M 2WC 33P 36B 3O- 4.4 476 53G 5GY 5RE 5VS AAHHS AANLZ AAQQT AASGY AAXRX AAZKR ABCUV ABEFU ABFTF ABJNI ABLJU ABQWH ABTAH ABXGK ACAHQ ACBEA ACCFJ ACCZN ACGFO ACGFS ACGOF ACPOU ACSMX ACXBN ACXQS ADBBV ADBTR ADKYN ADOZA ADXAS ADZMN AEEZP AEGXH AEIGN AENEX AEQDE AEUYR AFBPY AFFPM AHBTC AIACR AIAGR AIURR AIWBW AJBDE ALMA_UNASSIGNED_HOLDINGS ALUQN AMYDB ASPBG BFHJK C45 CS3 DCZOG DRFUL DRMAN DRSTM DU5 EBD EBS EJD EMB EMOBN F5P G8K HDBZQ HGLYW I-F KBYEO LATKE LEEKS LOXES LUTES LYRES MEWTI O9- OVD P2P P2W PALCI PHY RJQFR RNS ROL SAMSI SUPJJ SV3 TEORI TN5 TWZ USG WOHZO WXSBR XJT ZGI ZVN ZXP ZY4 ZZTAW AITYG AAMNL AAYXX ABDPE CITATION |
ID | FETCH-LOGICAL-c1774-ea468f12bc4fe505e63438eb40739a9cf683f9704fd0ff296c251fb669268bfe3 |
IEDL.DBID | 33P |
ISSN | 0094-2405 |
IngestDate | Fri Nov 22 00:52:12 EST 2024 Sat Aug 24 01:02:42 EDT 2024 Fri Jun 21 00:29:02 EDT 2024 |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 6 |
Language | English |
LinkModel | DirectLink |
MergedId | FETCHMERGED-LOGICAL-c1774-ea468f12bc4fe505e63438eb40739a9cf683f9704fd0ff296c251fb669268bfe3 |
PageCount | 1 |
ParticipantIDs | scitation_primary_10_1118_1_2961724 wiley_primary_10_1118_1_2961724_MP1724 crossref_primary_10_1118_1_2961724 |
PublicationCentury | 2000 |
PublicationDate | June 2008 |
PublicationDateYYYYMMDD | 2008-06-01 |
PublicationDate_xml | – month: 06 year: 2008 text: June 2008 |
PublicationDecade | 2000 |
PublicationTitle | Medical Physics |
PublicationYear | 2008 |
Publisher | American Association of Physicists in Medicine |
Publisher_xml | – name: American Association of Physicists in Medicine |
SSID | ssj0006350 |
Score | 1.9108081 |
Snippet | Purpose: In accelerated partial breast irradiation (APBI), accurate localization of the lumpectomy cavity is important. We are investigating daily localization... Abstract only Purpose: In accelerated partial breast irradiation (APBI), accurate localization of the lumpectomy cavity is important. We are investigating... |
SourceID | crossref wiley scitation |
SourceType | Aggregation Database Publisher |
StartPage | 2720 |
SubjectTerms | Anatomy Computed tomography Cone beam computed tomography Error analysis Image analysis Laser materials Medical imaging Radiography Skin Topology |
Title | SU‐GG‐J‐175: Target Registration Error Analysis Via KV Imaging and Conebeam CT in Accelerated Partial Breast Irradiation |
URI | http://dx.doi.org/10.1118/1.2961724 https://onlinelibrary.wiley.com/doi/abs/10.1118%2F1.2961724 |
Volume | 35 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://sdu.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV3JTsMwELWgEsuFpYAomyxA3CKaxnVsOJXSloKKKrqIWxQ7NuqhAaXtFfEJfCNfwthJI_UAQuKQ5GJHTsb2e-MZPyN0rqAfKKGF44ZMO4RJ5rBIg9cKcOtzpjlVdumi5z8-s9uGkcm5nu-FSfUh8gU3MzLsfG0GeCiyU0hck7gOzh038Gu0QMFLsNs3vG4-CwOQpttPODERhGqmKgR1L_OaC1i0BpCTRr8XeaoFmubmv5q4hTYyfolraYfYRksqLqLVThZBL6IVm_IpJzvovTf4-vhsteB2DxcA_BXu27xw_KRecj1d3EiS1wTPxUvwcBTihyFuj-3xRjiMI1x_jcFI4RjX-3gU45qUgGVGgiLCXdMzoT03Jvd9ittJYrQQzHt30aDZ6NfvnOw0Bke6wBEdFRLKtFsRkmgFvElRj3hMCWJifSGXmjJPc79MdFTWGj5cAnXSglJeoUxo5e2hQgzt2Ue4ShXhAqgL8BUgFBUuIh0B9xKeZIxX_RI6ndsleEtFN4LUWWGBG2T_tITOcov9VurCWujnEkGnax4Hfy14iNbTnBGzEnOECtNkpo7R8iSandge-A0_tdlq |
link.rule.ids | 310,311,315,782,786,791,792,1408,23939,23940,25149,27933,27934,46488 |
linkProvider | Wiley-Blackwell |
linkToHtml | http://sdu.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV3JTsMwEB2xCMqFHbFjAeIW0TauayMupXSjFFV0EbcodmzUQwMKcEV8At_IlzBO0kg9gJA4JLnYkZMZ-z17xs8Apxr9QEsjnYLPjUO54g4PDM5aEW7LghvBdLx00SvfPfDrmpXJuZzshUn0IbIFN9sz4vHadnC7IJ32cpu5jrM7YfGXzsI8ZeiIdgOH283GYYTSZAOKoDaGUEp1hbDyeVZ1Co1yCDpJ_HuaqcZQU1_5XyNXYTmlmKSS-MQazOhwHRY7aRB9HRbirE_1sgHvvcHXx2ejgbcbvBDjL0g_Tg0n9_oxk9QltSh6ishEv4QMRz5pD0lrHJ9wRPwwINWnEO3kj0m1T0YhqSiFcGZVKALStc6J7bmy6e-vpBVFVg7BvncTBvVav9p00gMZHFVAmuhonzJuCkWpqNFInTRzqcu1pDbc5wtlGHeNKOepCfLG4IcrZE9GMiaKjEuj3S2YC7E920BKTFMhkb0gZUFOURQyMAHSL-kqzkWpvAPHE8N4z4nuhpfMV7hX8NJ_ugMnmcl-K3UWm-jnEl6nax-7fy14BLlmv3Pr3bbu2nuwlKSQ2IWZfZh7jd70Acy-BG-HsTt-AxPl3ZI |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=proceeding&rft.title=Medical+Physics&rft.atitle=SU%E2%80%90GG%E2%80%90J%E2%80%90175%3A+Target+Registration+Error+Analysis+Via+KV+Imaging+and+Conebeam+CT+in+Accelerated+Partial+Breast+Irradiation&rft.au=Schuller%2C+B&rft.au=Nawaz%2C+A&rft.au=Wolfgang%2C+J&rft.au=Gierga%2C+D&rft.date=2008-06-01&rft.issn=0094-2405&rft.eissn=2473-4209&rft.volume=35&rft.issue=6&rft.spage=2720&rft.epage=2720&rft_id=info:doi/10.1118%2F1.2961724 |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0094-2405&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0094-2405&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0094-2405&client=summon |