21.3: NIST Colorimetric Calibration Facility for Displays - Part 2
Development of a calibration facility for spectroradiometric, colorimetric, and goniometric measurements of displays, display colorimeters, and spectroradiometers has been completed at the National Institute of Standards and Technology (NIST). An overview of the facility and results of the character...
Saved in:
Published in: | SID International Symposium Digest of technical papers Vol. 32; no. 1; pp. 330 - 333 |
---|---|
Main Authors: | , , |
Format: | Journal Article |
Language: | English |
Published: |
Oxford, UK
Blackwell Publishing Ltd
01-06-2001
|
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Development of a calibration facility for spectroradiometric, colorimetric, and goniometric measurements of displays, display colorimeters, and spectroradiometers has been completed at the National Institute of Standards and Technology (NIST). An overview of the facility and results of the characterization measurements are presented. The results show that the facility is capable of measuring colors of sources with an expanded uncertainty 0.001 for chromaticity (x,y) and 1 % for luminance (Y). The new calibration services from NIST are designed to reduce the uncertainty of color measurements using commercial instruments to a level approaching that of the NIST reference instrument. |
---|---|
Bibliography: | ark:/67375/WNG-VVJ4DQMT-R istex:69BE71D7841D1FAB8B47A036F31094095FE6A7FB ArticleID:SDTP3402 |
ISSN: | 0097-966X 2168-0159 |
DOI: | 10.1889/1.1831863 |