Effect of flow baffles on the dialysate flow distribution of hollow-fiber hemodialyzers: a nonintrusive experimental study using MRI
We used an innovative, nonintrusive MRI technique called the two-dimensional (2D) Phase-Contrast (2DPC) velocity-imaging technique to investigate the effect of flow baffles on the dialysate-side flow distribution in two different hollow-fiber hemodialyzers (A and B); each with flow rates between 200...
Saved in:
Published in: | Journal of biomechanical engineering Vol. 125; no. 4; p. 481 |
---|---|
Main Authors: | , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
United States
01-08-2003
|
Subjects: | |
Online Access: | Get more information |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We used an innovative, nonintrusive MRI technique called the two-dimensional (2D) Phase-Contrast (2DPC) velocity-imaging technique to investigate the effect of flow baffles on the dialysate-side flow distribution in two different hollow-fiber hemodialyzers (A and B); each with flow rates between 200 and 1000 mL/min (3.33 x 10(-6) and 1.67 x 10(-5) m3/s). Our experimental results show that (1) the dialysate-side flow distribution was nonuniform with channeling flow occurred at the peripheral cross section of these hollow-fiber hemodialyzers, and (2) the existing designs of flow baffles failed to promote uniform dialysate-side flow distribution for all flow rates studies. |
---|---|
ISSN: | 0148-0731 |
DOI: | 10.1115/1.1590355 |