Performance of the Abbott ID NOW rapid SARS-CoV-2 amplification assay in relation to nasopharyngeal viral RNA loads
•Calibration of the Xpert with viral stock dilutions was linear and reproducible.•Overall Xpert/ID-NOW positive agreement = 70% and negative agreement = 99.6%.•Sensitivity of ID-NOW for Xpert N2 CT ≤ 37 (viral loads ≥ 945 copies/mL) was 100%.•Sensitivity of ID-NOW for Xpert N2 for N2 CT > 37 was...
Saved in:
Published in: | Journal of clinical virology Vol. 140; p. 104843 |
---|---|
Main Authors: | , , , , , , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
Netherlands
Elsevier B.V
01-07-2021
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | •Calibration of the Xpert with viral stock dilutions was linear and reproducible.•Overall Xpert/ID-NOW positive agreement = 70% and negative agreement = 99.6%.•Sensitivity of ID-NOW for Xpert N2 CT ≤ 37 (viral loads ≥ 945 copies/mL) was 100%.•Sensitivity of ID-NOW for Xpert N2 for N2 CT > 37 was only 37.7%. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 1386-6532 1873-5967 1873-5967 |
DOI: | 10.1016/j.jcv.2021.104843 |