Significance of Nonuniform Attenuation Correction in Quantitative Brain SPECT Imaging

The purposes of this study were to develop a method for nonuniform attenuation correction of 123I emission brain images based on transmission imaging with a longer-lived isotope (i.e., 57Co) and to evaluate the relative improvement in quantitative SPECT images achieved with nonuniform attenuation co...

Full description

Saved in:
Bibliographic Details
Published in:The Journal of nuclear medicine (1978) Vol. 39; no. 10; pp. 1719 - 1726
Main Authors: Rajeevan, Nallakkandi, Zubal, I. George, Ramsby, S. Quinn, Zoghbi, Sami S, Seibyl, John, Innis, Robert B
Format: Journal Article
Language:English
Published: Reston, VA Soc Nuclear Med 01-10-1998
Society of Nuclear Medicine
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The purposes of this study were to develop a method for nonuniform attenuation correction of 123I emission brain images based on transmission imaging with a longer-lived isotope (i.e., 57Co) and to evaluate the relative improvement in quantitative SPECT images achieved with nonuniform attenuation correction. Emission and transmission SPECT scans were acquired on three different sets of studies: a heterogeneous brain phantom filled with 1231 to simulate the distribution of dopamine transporters labeled with 2beta-carbomethoxy-3beta-(4-123I-iodophenyl)tropane (123I-beta-CIT); nine healthy human control subjects who underwent transmission scanning using two separate line sources (57Co and 123I); and a set of eight patients with Parkinson's disease and five healthy control subjects who received both emission and transmission scans after injection of 123I-beta-CIT. Attenuation maps were reconstructed using a Bayesian transmission reconstruction algorithm, and attenuation correction was performed using Chang's postprocessing method. The spatial distribution of errors within the brain was obtained from attenuation correction factors computed from uniform and nonuniform attenuation maps and was visualized on a pixel-by-pixel basis as an error image. For the heterogeneous brain phantom, the uniform attenuation correction had errors of 2%-6.5% for regions corresponding to striatum and background, whereas nonuniform attenuation correction was within 1%. Analysis of 123I transmission images of the nine healthy human control subjects showed differences between uniform and nonuniform attenuation correction to be in the range of 6.4%-16.0% for brain regions of interest (ROIs). The human control subjects who received transmission scans only were used to generate a curvilinear function to convert 57Co attenuation values into those for 123I, based on a pixel-by-pixel comparison of two coregistered transmission images for each subject. These values were applied to the group of patients and healthy control subjects who received transmission 57Co scans and emission 123I scans after injection of 123I-beta-CIT. In comparison to nonuniform attenuation correction as the gold standard, uniform attenuation with the ellipse drawn around the transmission image caused an approximately 5% error, whereas placement of the ellipse around the emission image caused a 15% error. Nonuniform attenuation correction allowed a moderate improvement in the measurement of absolute activity in individual brain ROIs. When images were analyzed as target-to-background activity ratios, as is commonly performed with 123I-beta-CIT, these outcome measures showed only small differences when Parkinson's disease patients and healthy control subjects were compared using nonuniform, uniform or even no attenuation correction.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0161-5505
1535-5667