CALIBRATION OF GAFCHROMIC EBT3 FILM FOR DOSIMETRY OF SCANNING PROTON PENCIL BEAM (PBS)
Gafchromic EBT3 films are applied in proton radiotherapy for 2D dose mapping because they demonstrate spatial resolution well below 1 mm. However, the film response must be corrected in order to reach the accuracy of dose measurements required for the clinical use. The in-house developed AnalyseGafc...
Saved in:
Published in: | Radiation protection dosimetry Vol. 180; no. 1-4; pp. 324 - 328 |
---|---|
Main Authors: | , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
England
01-08-2018
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Gafchromic EBT3 films are applied in proton radiotherapy for 2D dose mapping because they demonstrate spatial resolution well below 1 mm. However, the film response must be corrected in order to reach the accuracy of dose measurements required for the clinical use. The in-house developed AnalyseGafchromic software allows to analyze and correct the measured response using triple channel dose calibration, statistical scan-to-scan fluctuations as well as experimentally determined dose and LET dependence. Finally, the optimized protocol for evaluation of response of Gafchromic EBT3 films was applied to determine 30 × 40 cm2 dose profiles of the scanning therapy unit at the Cyclotron Centre Bronowice, CCB in Krakow, Poland. |
---|---|