SU-F-T-622: Comparative Analysis of Pencil Beam and Anisotropic Analytical Algorithm (AAA) for Stereotactic Body Radiation Therapy (SBRT) of Thoracic Spine
Purpose: To compare the impact of Pencil Beam(PB) and Anisotropic Analytic Algorithm(AAA) dose calculation algorithms on OARs and planning target volume (PTV) in thoracic spine stereotactic body radiation therapy (SBRT). Methods: Ten Spine SBRT patients were planned on Brainlab iPlan system using hy...
Saved in:
Published in: | Medical physics (Lancaster) Vol. 43; no. 6; p. 3607 |
---|---|
Main Authors: | , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
United States
American Association of Physicists in Medicine
01-06-2016
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Purpose:
To compare the impact of Pencil Beam(PB) and Anisotropic Analytic Algorithm(AAA) dose calculation algorithms on OARs and planning target volume (PTV) in thoracic spine stereotactic body radiation therapy (SBRT).
Methods:
Ten Spine SBRT patients were planned on Brainlab iPlan system using hybrid plan consisting of 1–2 non-coplanar conformal-dynamic arcs and few IMRT beams treated on NovalisTx with 6MV photon. Dose prescription varied from 20Gy to 30Gy in 5 fractions depending on the situation of the patient. PB plans were retrospectively recalculated using the Varian Eclipse with AAA algorithm using same MUs, MLC pattern and grid size(3mm).Differences in dose volume parameters for PTV, spinal cord, lung, and esophagus were analyzed and compared for PB and AAA algorithms. OAR constrains were followed per RTOG-0631.
Results:
Since patients were treated using PB calculation, we compared all the AAA DVH values with respect to PB plan values as standard, although AAA predicts the dose more accurately than PB. PTV(min), PTV(Max), PTV(mean), PTV(D99%), PTV(D90%) were overestimated with AAA calculation on average by 3.5%, 1.84%, 0.95%, 3.98% and 1.55% respectively as compared to PB. All lung DVH parameters were underestimated with AAA algorithm mean deviation of lung V20, V10, V5, and 1000cc were 42.81%,19.83%, 18.79%, and 18.35% respectively. AAA overestimated Cord(0.35cc) by mean of 17.3%; cord (0.03cc) by 12.19% and cord(max) by 10.5% as compared to PB. Esophagus max dose were overestimated by 4.4% and 5cc by 3.26% for AAA algorithm as compared to PB.
Conclusion:
AAA overestimated the PTV dose values by up to 4%.The lung DVH had the greatest underestimation of dose by AAA versus PB. Spinal cord dose was overestimated by AAA versus PB. Given the critical importance of accuracy of OAR and PTV dose calculation for SBRT spine, more accurate algorithms and validation of calculated doses in phantom models are indicated. |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1118/1.4956807 |