Direct Gating of Microchannel Plates

We investigated whether direct fast gating of microchannel plates (MCP) is possible. Our experiments measured the MCP gain as a function of time (on the nanosecond time scale) for an MCP intensifier tube that was gated by applying voltage to the plate itself. We also examined the dynamic range of an...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on nuclear science Vol. 30; no. 6; pp. 4624 - 4627
Main Authors: Jacoby, Barry A., Kotecki, David E., Lear, Richard D.
Format: Journal Article
Language:English
Published: IEEE 01-01-1983
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We investigated whether direct fast gating of microchannel plates (MCP) is possible. Our experiments measured the MCP gain as a function of time (on the nanosecond time scale) for an MCP intensifier tube that was gated by applying voltage to the plate itself. We also examined the dynamic range of an MCP that was being gated. Our results showed that, though an MCP takes milliseconds to recharge, fast gating is not incompatible with this slow recharge time.
ISSN:0018-9499
1558-1578
DOI:10.1109/TNS.1983.4333185