Simulating NEMA characteristics of the modular total-body J-PET scanner -- an economic total-body PET from plastic scintillators
The purpose of the presented research is estimation of the performance characteristics of the economic Total-Body Jagiellonian-PET system (TB-J-PET) constructed from plastic scintillators. The characteristics are estimated according to the NEMA NU-2-2018 standards utilizing the GATE package. The sim...
Saved in:
Main Authors: | , , , , , , , , , , , , , , , , , , , , , , , , , , , , |
---|---|
Format: | Journal Article |
Language: | English |
Published: |
03-07-2021
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The purpose of the presented research is estimation of the performance
characteristics of the economic Total-Body Jagiellonian-PET system (TB-J-PET)
constructed from plastic scintillators. The characteristics are estimated
according to the NEMA NU-2-2018 standards utilizing the GATE package. The
simulated detector consists of 24 modules, each built out of 32 plastic
scintillator strips (each with cross section of 6 mm times 30 mm and length of
140 cm or 200 cm) arranged in two layers in regular 24-sided polygon
circumscribing a circle with the diameter of 78.6 cm. For the TB-J-PET with an
axial field-of-view (AFOV) of 200 cm, a spatial resolutions of 3.7 mm
(transversal) and 4.9 mm (axial) are achieved. The NECR peak of 630 kcps is
expected at 30 kBq/cc activity concentration and the sensitivity at the center
amounts to 38 cps/kBq. The SF is estimated to 36.2 %. The values of SF and
spatial resolution are comparable to those obtained for the state-of-the-art
clinical PET scanners and the first total-body tomographs: uExplorer and
PennPET. With respect to the standard PET systems with AFOV in the range from
16 cm to 26 cm, the TB-J-PET is characterized by an increase in NECR
approximately by factor of 4 and by the increase of the whole-body sensitivity
by factor of 12.6 to 38. The TOF resolution for the TB-J-PET is expected to be
at the level of CRT=240 ps (FWHM). For the TB-J-PET with an axial field-of-view
(AFOV) of 140 cm, an image quality of the reconstructed images of a NEMA IEC
phantom was presented with a contrast recovery coefficient (CRC) and a
background variability parameters. The increase of the whole-body sensitivity
and NECR estimated for the TB-J-PET with respect to current commercial PET
systems makes the TB-J-PET a promising cost-effective solution for the broad
clinical applications of total-body PET scanners. |
---|---|
DOI: | 10.48550/arxiv.2107.01356 |