Development of a medical fiber-optic oxygen sensor based on optical absorption change

A fiber-optic oxygen sensor has been developed for use in medical applications. The sensor's viologen indicator becomes strongly absorbant after brief UV stimulation, and then returns to the transparent state. The rate of indicator return to transparency is proportional to the local oxygen conc...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on biomedical engineering Vol. 39; no. 2; pp. 185 - 193
Main Authors: Wolthius, R., McCrae, D., Hartl, J.C., Saaski, E., Mitchell, G.L., Garcin, K., Willard, R.
Format: Journal Article
Language:English
Published: United States IEEE 01-02-1992
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A fiber-optic oxygen sensor has been developed for use in medical applications. The sensor's viologen indicator becomes strongly absorbant after brief UV stimulation, and then returns to the transparent state. The rate of indicator return to transparency is proportional to the local oxygen concentration. Indicator absorbance is monitored with a red LED and receiving photodiode, and absorbance data are processed by a dedicated CPU. The solid-state sensor system has performance comparable to existing oxygen measurement techniques, and may be applicable for both in vitro and in vivo oxygen measurements.< >
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
ObjectType-Article-1
ObjectType-Feature-2
ISSN:0018-9294
1558-2531
DOI:10.1109/10.121650