Comparison of white-to-white measurements using four devices and their determination of ICL sizing
Background To compare the measurements obtained from the Orbscan II, IOLMaster 700, Pentacam AXL, and Castroviejo caliper and their effects on calculating the recommended implantable collamer lens (ICL) size and postoperative vault measurements. Methods This is a retrospective cross-sectional study...
Saved in:
Published in: | Eye and vision (Novato, Calif.) Vol. 9; no. 1; pp. 1 - 36 |
---|---|
Main Authors: | , , , , |
Format: | Journal Article |
Language: | English |
Published: |
London
BioMed Central Ltd
02-10-2022
BioMed Central BMC |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background To compare the measurements obtained from the Orbscan II, IOLMaster 700, Pentacam AXL, and Castroviejo caliper and their effects on calculating the recommended implantable collamer lens (ICL) size and postoperative vault measurements. Methods This is a retrospective cross-sectional study of patients who underwent ICL surgery by a single surgeon from March 1, 2018 to July 31, 2021. Records were reviewed for the anterior chamber depth (ACD) and white-to-white (WTW) measurements obtained from the Orbscan II, IOLMaster 700, Pentacam AXL, and Castroviejo caliper (WTW only). These were used to calculate the recommended ICL size. The actual ICL size implanted, and vault measurements obtained one month postoperatively were also collected. Results One hundred seven eyes with a mean age of 27.9 [+ or -] 7.7 years were included in the study. Mean WTW measurements were significantly different between devices (P < 0.0001), with the IOLMaster 700 having the highest value (12.14 [+ or -] 0.04 mm) and the caliper having the lowest value (11.45 [+ or -] 0.04 mm). Mean ACD measurements were the lowest in Orbscan II (3.12 [+ or -] 0.25 mm) and the highest in Pentacam AXL (3.16 [+ or -] 0.24 mm). The Pentacam AXL produced an ICL size similar to the Orbscan in 69.2% of eyes. The IOLMaster yielded an ICL measurement one size larger than Orbscan-based calculations in 64.5% of eyes. Using the Orbscan WTW and ACD, the desired vault of 0.25 to 0.75 mm and 0.25 to 1.00 mm was achieved in 70% and 91% of eyes, respectively. Substituting caliper WTW to IOLMaster 700 or Pentacam AXL WTW increases the percentage of achieving the desired vault to 80%, similar to the Orbscan. Conclusions The Orbscan II, IOLMaster 700, and Pentacam AXL cannot be used interchangeably for calculating ICL sizing. Combining the WTW from caliper measurement with the ACD of the IOLMaster 700 or Pentacam AXL could improve ICL sizing and achieve a higher percentage of eyes with the desired vault. Keywords: ICL vault, Orbscan II, IOLMaster 700, Pentacam AXL, Caliper |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 2326-0254 2326-0246 2326-0254 |
DOI: | 10.1186/s40662-022-00308-z |